The high temperature insertion type electrode is designed for periodic exposure to steam sterilization or continuous high temperature. This pH electrode is housed in a durable thermoplastic (PAS) body and is rated for temperatures up to 135°C. The use of the double Porous PTFE liquid junctions with matched viscosity electrolytes provides a reference cell which permits extended periods of pH measurements in the presence of sulfides or other silver complexing agents. This electrode also incorporates pressure compensating devices to protect the electrode from extreme temperature induced pressure deviations. Ten foot cable length and BNC connector are standard.
Application • High Temperature Environment • Continuous Processing Applications • Harsh Conditions • Steam Sterilization
Specifications pH Range: 0 to 14 pH Temperature Range: -5 to +135°C (23 to 275°F) Pressure Range: 500 psig @ 25°C Accuracy: ±0.1% over full range Impedance: 125 Mohm (standard version) Reference Cell: Double Junction, KCl/AgCl, KNO3 Reference Junction: Porous PTFE Wetted Materials: PTFE, Glass membrane, polymer outer body, EPR “O” Rings Response Time: 95% of reading in 10 seconds Drift: Less than 2 mV per week