The CL110 is a precision loop calibrator that features a user friendly interface with dedicated buttons and a quick set knob. This combination dramatically reduces the time it takes to measure or source current and power up a loop. The rugged case is contoured to fit a technicians hand and the large backlit graphics LCD is easy-to-read in any lighting.
SPECIFICATIONS Functions: mA source, simulate, read or read with loop power, voltage read only Ranges: 0 to 24 mA, 0 to 30 Vdc (read only) Resolution: 1 uA on mA range and 1 mV on voltage range Accuracy: ±0.01% rdg +2 LSD all ranges (@23 ±5°C) Operating Temperature: -10 to 55°C (14 to 131°F) Temperature Effect: 20 ppm/°C from -10 to 18°C and 28 to 55°C Humidity Range: 10 to 95% non-condensing Display: 128 x 64 pixel LCD graphic display with backlight, 9 mm (0.34') high digits Power: 6 “AAA” batteries (alkaline, lithium or NiMH) included Battery Life: > 40 hours continuous (measure mode using alkaline) Loop Compliance Voltage: 24 Vdc @ 20 mA Over-Voltage Protection: 240 Vac Overload Current Protection: 28 mA DC Dimensions: 150 L x 90 W x 30 mm D (6 x 3.6 x 1.3') Weight: 0.3 kg (9.5 oz)
Hello,
Do any of these loop calibrators have CSA or CUL approval ratings ??
Asked by: Onyx202
CL110 does not have CSA or CUL Approvals.
Answered by: Applicatrions Engineer JK
Date published: 2021-04-07
In mA Source mode, with nothing attached to it, the display will alternate showing output level and "OL". I cannot find this in the manual. What does this mean?
Asked by: Islandan
Hi Islandan,
Thank you for the inquiry. This calibrator can provide signal to resitive loads up to 1200Ω without HART resistor, 950 Ω with HART resistor. Reistances greater than these or open line result in error message "OL".
Best regards,